Test Code Mapping for GeneXpert Xpress_SARS-CoV-2_Flu_RSV ‘COVID Quad’ Panel

Trucos y Sugerencias

This guide describes the test code mapping required to report the interpreted results (POS/NEG) for the SARSCOV2, FLUA, FLUB, and RSV tests, as well as how to send an order for the panel from your LIS/EMR. This document assumes that the Cepheid GeneXpert is configured with the default codes as shown here:

ADF Name / Assay NameADF VersionSuggested Host Order Codeb(Assay Host Test Code)Analyte Common NameSuggested GeneXpert Result Code
Xpert Xpress_SARS-CoV-2_Flu_RSV3SARSCOV2FLURSVSARS-CoV-2SARSCOV2
FLU AFLUA
FLU BFLUB
RSVRSV

For information on configuring the GeneXpert codes, see the Operator’s Manual or contact Cepheid Support.

Driver Configuration

The first step is to open Driver Properties and check Parse Interpreted Results as separate Test Code. If this is not enabled, the interpreted results will not be parsed as a test result value.

Result Code Mapping

A SARSCOV2FLURSV result message from the GeneXpert contains 31 observations, but you should only map the tests that you will be reporting. This section explains how to map the GeneXpert result codes so that you can report the interpreted results for SARSCOV2, FLUA, FLUB, and RSV.

Analyte Common NameInstrument Test Code
SARS-CoV-2SARSCOV2-Xpert Xpress_SARS-CoV-2_Flu_RSV-SARS-CoV-2-I
FLU AFLUA-Xpert Xpress_SARS-CoV-2_Flu_RSV-FLU A-I
FLU BFLUB-Xpert Xpress_SARS-CoV-2_Flu_RSV-FLU B-I
RSVRSV-Xpert Xpress_SARS-CoV-2_Flu_RSV-RSV-I

Open the Test Code Mapping window for your GeneXpert configuration and select Add. For Instrument Code enter “SARSCOV2-Xpert Xpress_SARS-CoV-2_Flu_RSV-SARS-CoV-2-I” and enter a description for the Covid test code. Select Add and then Close.

Select the newly created “SARSCOV2-Xpert Xpress_SARS-CoV-2_Flu_RSV-SARS-CoV-2-I” in the Select Instrument Test Code field. In the lower Results mapping section, select the existing Instrument Test Code, then select Remove.

In the Select or Add IM Test Code field, enter the test code for Covid used by your LIS/EMR. Change Select Test Resulting Option to Default (D). Select Add. Your screen should look like this:

Repeat the above steps to add the other three instrument test codes.

Order Code Mapping

This section explains how to add an order mapping so that you can send an order for the Xpert Xpress_SARS-CoV-2_Flu_RSV assay to the GeneXpert. It is assumed that you completed the Result Code Mapping section above.

Open the Test Code Mapping window for your GeneXpert configuration and select Add. For Instrument Code enter “SARSCOV2FLURSV” and enter a description for the COVID panel code. Select Add and then Close.

Select the newly created “SARSCOV2FLURSV” in the Select Instrument Test Code field. In the upper Orders mapping section, select the existing Instrument Test Code, then select Remove.

In the Select or Add IM Test Code box, enter the test code for Covid used by your LIS/EMR. Change ‘Select Fluid Type’ to “Other”. Click Add. Your screen should look like this:

More Tips & Tricks Articles

Trucos y Sugerencias

TCP/IP Keep-Alive

Data Innovations’ (DI’s) Instrument Manager (IM) offers connectivity that enables integration with virtually any instrument, LIS, or EMR/EHR regardless of...

Instrument Manager, Tips & Tricks

Following Driver Updates in My DI Community

Descripción general

Data Innovations (DI) provides ways to stay up to date on the latest driver updates.The first way is right through...
Instrument Manager, Tips & Tricks

Net Gateway Startup Issues (Error 5023)

Descripción general

When IRIS and Instrument Manager (IM) are starting after a server shuts down or restarts, the .NET gateway is not...
Trucos y Sugerencias

How to Install Instrument Manager™ Core Software

NOTE These instructions apply for Data Innovations’ North American customers. If this is your first time installing IM, please contact Data...

Trucos y Sugerencias

How to help Data Innovations help you when requesting Support

Data Innovations' Tips & Tricks are intended to promote the effective and productive use of Instrument Manager. Fortunately, Instrument Manager...

Trucos y Sugerencias

Evaluation Rules for Clinical Ranges Precaution

When building an Autoverification Rule Set it's common to include evaluation rules for Clinical Ranges. These rules typically set test...

Trucos y Sugerencias

Instrument Manager System Preventative Maintenance (volume 1)

Instrument Manager is a robust software application with a proven track record of operational stability. However, as with any software...

Trucos y Sugerencias

Instrument Manager System Preventative Maintenance (volume 2)

This month's Tips & Tricks is the second installation of Preventative Maintenance recommendations intended to help you maintain a stable...

¿Quiere conocer más?